QUASAR™ IsoCenter Cube Phantom

The QUASAR™ IsoCenter Cube is designed to implement a modified Winston-Lutz test for IGRT accuracy required for optimal treatment delivery. This system is also used for TG-142 quality assurance compliance for isocenter and laser alignment tests.

End-to-end positioning accuracy from planning CT to clinical IGRT treatment setup is achieved with 0.1 mm precision and is accomplished in minutes. The test is performed by aligning the QUASAR™ IsoCenter Cube phantom using the cross-hairs etched on each side with your IGRT system, then capturing radiation field exposures with film or EPID. The test results are used to determine end-to-end isocenter accuracy for frameless SRS, SBRT, and IGRT treatments.

  • Specifications

    • Acrylic cube; 5 x 5 x 5 cm
    • Tungsten sphere at isocenter 6mm diameter
    • Rubber no-slip base
    • Laser alignment marks on cube
    • Offset laser alignment marks on 3 faces
  • Minimum Technical Requirements

    • Standard PC desktop or laptop with Windows 7, Vista or XP
    • 1 GB RAM
    • .NET framework 4.0 for client and server
  • 100-1016: QUASAR™ IsoCenter Cube Package, includes:

    • Phantom
    • Software
  • 100-1016-100: IsoCenter Cube Phantom

    • Additional Phantom
Download Product Data Sheet

Features

Workflow Efficiency

Quick and simple Winston-Lutz testing for IGRT accuracy. Integrate the QUASAR™ IsoCentre Cube into your daily QA routine for simple verification of your laser, isocentre and couch adjustment alignment.

Simple Design

The QUASAR™ IsoCenter Cube Phantom features a seamless 5 cm³ acrylic unibody enclosure containing one 6 mm diameter Tungsten Sphere located at the central axis of the main cross-hairs and used for measuring isocenter accuracy.

Image Match Verification

Displayed on three faces of the QUASAR™ IsoCenter Cube Phantom are 3 mm offset laser alignment marks, which are used to easily verify the couch position adjustment accuracy after image matching.

Isocentre Verification

User confidence is increased with the verification of your isocentre alignment accuracy. Report detailed statistics including left/right, superior/inferior and anterior/posterior errors at configurable gantry, collimator and table angles.

TG142 Compliance

Perform required testing to confirm the accuracy and precision of your isocenter alignment. End-to-end positioning accuracy from planning CT to clinical IGRT treatment setup is achieved with 0.1 mm precision and is accomplished in minutes.

Software

Automated software is included with the Isocentre Cube Phantom to quickly analyze isocenter accuracy and EPID exposure quality. Confirm your visual imaging analysis with statistical data and user-selected pass/fail criteria.

Exposure Evaluation

Analyze EPID radiation field exposures with the automated Analysis Software. Check various beam shaping devices, report left/right, superior/inferior and anterior/posterior errors at configurable gantry, collimator and table angles.

Intuitive Analysis

Ability to test IGRT accuracy with a minimum of two gantry angle images. The user sets the warning and fail levels, and can visually see pass (green), warning (yellow), and fail (red) results.

Product Information & Accessories